Configuration of the external trigger sources. Threshold is ignored for entries 8 and above.
The trigger indexes refer to the entry in the trigger array and are defined like this:
#define NDIGO TRIGGER A0 0
#define NDIGO TRIGGER A1 1
#define NDIGO TRIGGER B0 2
#define NDIGO TRIGGER B1 3
#define NDIGO TRIGGER C0 4
#define NDIGO TRIGGER C1 5
#define NDIGO TRIGGER D0 6
#define NDIGO TRIGGER D1 7
#define NDIGO TRIGGER TDC 8
#define NDIGO TRIGGER GATE 9
#define NDIGO TRIGGER BUS0 10
#define NDIGO TRIGGER BUS1 11
#define NDIGO TRIGGER BUS2 12
#define NDIGO TRIGGER BUS3 13
#define NDIGO TRIGGER AUTO 14
#define NDIGO TRIGGER ONE 15
#define NDIGO TRIGGER TDC PE 16
#define NDIGO TRIGGER GATE PE 17
#define NDIGO TRIGGER BUS0 PE 18
#define NDIGO TRIGGER BUS1 PE 19
#define NDIGO TRIGGER BUS2 PE 20
#define NDIGO TRIGGER BUS3 PE 21
Sources 16 to 21 always are positive edge triggered.
ndigo trigger block trigger block[NDIGO CHANNEL COUNT + 1]
A structure describing the trigger settings of the four channels plus the timestamp channel.
ndigo gating block gating block[NDIGO250M GATE COUNT]
A structure describing the gating blocks that can be used by the trigger blocks to filter triggers.
int drive bus[4]
Enable output drive for each of the four external sync lines. Each integer represents a bitmask
selecting the trigger sources for that line. The bit mapping is described in section “Structure
ndigo trigger block on page 35.
int auto trigger period
int auto trigger random exponent
Create a trigger either periodically or randomly. There are two parameters M = trigger period
and N = random exponent that result in a distance between triggers of