#define NDIGO TRIGGER SOURCE GATE 0x00000200
#define NDIGO TRIGGER SOURCE BUS0 0x00000400
#define NDIGO TRIGGER SOURCE BUS1 0x00000800
#define NDIGO TRIGGER SOURCE BUS2 0x00001000
#define NDIGO TRIGGER SOURCE BUS3 0x00002000
#define NDIGO TRIGGER SOURCE AUTO 0x00004000
#define NDIGO TRIGGER SOURCE ONE 0x00008000
#define NDIGO TRIGGER SOURCE TDC PE 0x01000000
#define NDIGO TRIGGER SOURCE GATE PE 0x02000000
#define NDIGO TRIGGER SOURCE BUS0 PE 0x04000000
#define NDIGO TRIGGER SOURCE BUS1 PE 0x08000000
#define NDIGO TRIGGER SOURCE BUS2 PE 0x10000000
#define NDIGO TRIGGER SOURCE BUS3 PE 0x20000000
int gates
#define NDIGO TRIGGER GATE NONE 0x0000
#define NDIGO TRIGGER GATE 0 0x0001
#define NDIGO TRIGGER GATE 1 0x0002
#define NDIGO TRIGGER GATE 2 0x0004
#define NDIGO TRIGGER GATE 3 0x0008
double minimum free packets;
This parameter sets how many packets are supposed to fit into the on-board FIFO before a new
packet is started after the FIFO was full, i.e. a certain amount of free space in the FIFO is
demanded before a new packet is written after the FIFO was full. As a measure for the packet
length the gatelength set by the user is used. The on-board algorithm checks the free FIFO
space only in case the FIFO is full. Therefore, if this number is 1.0 or more at least every second
packet in the DMA buffer is guaranteed to have the full length set by the gatelength parameters.
In many cases smaller values will also result in full length packets. But below a certain value
multiple packets that are cut off at the end will show up.
3.4.4 Structure ndigo gating block
ndigo bool t negate
Invert output polarity. Defaults to false.
ndigo bool t retrigger
Defaults to false. If retriggering is enabled the timer is reset to the value of the start parameter
whenever the input signal is set while waiting to reach the stop time.
ndigo bool t extend
Defaults to true. If set, a gate is created with the set timing from the first occurrence of the
input trigger even for short gates. If not set, the input signal must persist for the gate to be
cronologic GmbH & Co. KG 36 Ndigo250M-14 User Guide